Dynamomètre multicomposant MicroDyn, 500 N, plaque supérieure 30x30 mm / 9109AA

Dynamomètre multicomposant MicroDyn, 500 N, plaque supérieure 30x30 mm
9109AA
18034487
  • Dynamomètre piézoélectrique à plusieurs composants avec forces maximales Fx, Fy, Fz ±500 N
  • Plaque supérieure 30x30 mm
  • Fréquence propre fn (x), fn (y), fn (z) >15 kHz
  • Connecteur à bride Fischer à 9 broches
Sur demande
Sur demande
Measuring ranges
Minimum measuring range [kN]
-0.5 kN
Maximum measuring range [kN]
0.5 kN
Calibrations
Calibrated range Fx [kN]
0 … 0.5 kN
Calibrated range Fy [kN]
0 … 0.5 kN
Calibrated range Fz [kN]
0 … 0.5 kN
Product type
Force sensor type
A dynamometer is a force plate measuring forces and moments.
Multicomponent dynamometer
Preloaded sensor
Preloaded sensors are already calibrated and therefore ready-to-use. Non-preloaded sensors are smaller.
Yes
Mechanical properties
Overload Fx [kN]
-1, 1 kN
Overload Fy [kN]
-1, 1 kN
Overload Fz [kN]
-1, 1 kN
Overload Mx [N·m]
-30, 30 N·m
Overload My [N·m]
-30, 30 N·m
Overload Mz [N·m]
-60, 60 N·m
Crosstalk
Maximum crosstalk Fx→Fy
-4 … 4 %
Maximum crosstalk Fx→Fz
-2 … 2 %
Maximum crosstalk Fy→Fx
-4 … 4 %
Maximum crosstalk Fy→Fz
-2 … 2 %
Maximum crosstalk Fz→Fx
-2 … 2 %
Maximum crosstalk Fz→Fy
-2 … 2 %
Metrological characteristics
Linearity (force)
Linearity represents the maximum deviation/error between ideal and actual output signal characteristics in relation to the measurand in a specific measuring range. It is expressed in percentage of the range of measurement signal (Full Scale Output).
≤±0.5 %/FSO
Hysteresis (force)
The maximum difference in output, at any measurand value within the specified range, when the value is approached first increasing and then decreasing measurand.
≤±0.5 %/FSO
Nominal sensitivity [pC/N]
-20 pC/N
Nominal sensitivity Fx [pC/N]
-12.5 pC/N
Nominal sensitivity Fy [pC/N]
-12.5 pC/N
Threshold (force)
Minimum change in the measurand that produces a measurable change in the sensor output, while the change of the measurand takes place slowly and monotonically.
0.002 N
Electrical properties
Output signal type
Charge (Piezoelectric, PE)
Capacitance
≈240 pF
Minimal insulation resistance
10^13 Ω
Ground Insulation
>10^8 Ω
Influence quantities
Sensitivity
Fx: -12.5 pC/N, Fy: -12.5 pC/N, Fz: -20 pC/N
Dynamical properties
Natural frequency Fx
Frequency of free (not forced) oscillations of the entire sensor. In practice the (usually lower) natural frequency of the entire mounted system governs the frequency behavior.
15 kHz
Natural frequency Fy
Frequency of free (not forced) oscillations of the entire sensor. In practice the (usually lower) natural frequency of the entire mounted system governs the frequency behavior.
15 kHz
Natural frequency Fz
Frequency of free (not forced) oscillations of the entire sensor. In practice the (usually lower) natural frequency of the entire mounted system governs the frequency behavior.
15 kHz
Certificates and Standards
Degree of protection EN 60529
IP67
Operation and installation
Operating temperature range
-20 … 70 °C
Applications
Cutting processes and machining development
Cable properties
Cable included
No cable included
Interfaces
Connector type
Fischer 104 9-pin (negative)
Dimensions and materials
Weight
1.04 kg
Il s'agit d'un dynamomètre multicomposantes permettant de mesurer les trois composantes orthogonales d'une force et le couple Mz. Le seuil très bas et la sensibilité élevée permettent de mesurer même les forces les plus faibles. Avec une fréquence propre supérieure à 15 kHz dans les trois composantes, des forces hautement dynamiques peuvent être enregistrées avec une très grande précision.
Chargement
Chargement