Dinamómetro Multicomponente MicroDyn, 500 N, Cubierta 30x30 mm / 9109AA

Dinamómetro Multicomponente MicroDyn, 500 N, Cubierta 30x30 mm
9109AA
18034487
  • Dinamómetro multicomponente piezoeléctrico con fuerzas máximas Fx, Fy, Fz ±500 N
  • Cubierta 30x30 mm
  • Frecuencia natural fn (x), fn (y), fn (z) >15 kHz
  • Conector Fischer de brida de 9 polos
Bajo pedido
Bajo pedido
Measuring ranges
Minimum measuring range [kN]
-0.5 kN
Maximum measuring range [kN]
0.5 kN
Calibrations
Calibrated range Fx [kN]
0 … 0.5 kN
Calibrated range Fy [kN]
0 … 0.5 kN
Calibrated range Fz [kN]
0 … 0.5 kN
Product type
Force sensor type
A dynamometer is a force plate measuring forces and moments.
Multicomponent dynamometer
Preloaded sensor
Preloaded sensors are already calibrated and therefore ready-to-use. Non-preloaded sensors are smaller.
Yes
Mechanical properties
Overload Fx [kN]
-1, 1 kN
Overload Fy [kN]
-1, 1 kN
Overload Fz [kN]
-1, 1 kN
Overload Mx [N·m]
-30, 30 N·m
Overload My [N·m]
-30, 30 N·m
Overload Mz [N·m]
-60, 60 N·m
Crosstalk
Maximum crosstalk Fx→Fy
-4 … 4 %
Maximum crosstalk Fx→Fz
-2 … 2 %
Maximum crosstalk Fy→Fx
-4 … 4 %
Maximum crosstalk Fy→Fz
-2 … 2 %
Maximum crosstalk Fz→Fx
-2 … 2 %
Maximum crosstalk Fz→Fy
-2 … 2 %
Metrological characteristics
Linearity (force)
Linearity represents the maximum deviation/error between ideal and actual output signal characteristics in relation to the measurand in a specific measuring range. It is expressed in percentage of the range of measurement signal (Full Scale Output).
≤±0.5 %/FSO
Hysteresis (force)
The maximum difference in output, at any measurand value within the specified range, when the value is approached first increasing and then decreasing measurand.
≤±0.5 %/FSO
Nominal sensitivity [pC/N]
-20 pC/N
Nominal sensitivity Fx [pC/N]
-12.5 pC/N
Nominal sensitivity Fy [pC/N]
-12.5 pC/N
Threshold (force)
Minimum change in the measurand that produces a measurable change in the sensor output, while the change of the measurand takes place slowly and monotonically.
0.002 N
Electrical properties
Output signal type
Charge (Piezoelectric, PE)
Capacitance
≈240 pF
Minimal insulation resistance
10^13 Ω
Ground Insulation
>10^8 Ω
Influence quantities
Sensitivity
Fx: -12.5 pC/N, Fy: -12.5 pC/N, Fz: -20 pC/N
Dynamical properties
Natural frequency Fx
Frequency of free (not forced) oscillations of the entire sensor. In practice the (usually lower) natural frequency of the entire mounted system governs the frequency behavior.
15 kHz
Natural frequency Fy
Frequency of free (not forced) oscillations of the entire sensor. In practice the (usually lower) natural frequency of the entire mounted system governs the frequency behavior.
15 kHz
Natural frequency Fz
Frequency of free (not forced) oscillations of the entire sensor. In practice the (usually lower) natural frequency of the entire mounted system governs the frequency behavior.
15 kHz
Certificates and Standards
Degree of protection EN 60529
IP67
Operation and installation
Operating temperature range
-20 … 70 °C
Applications
Cutting processes and machining development
Cable properties
Cable included
No cable included
Interfaces
Connector type
Fischer 104 9-pin negative
Dimensions and materials
Weight
1.04 kg
Se trata de un dinamómetro multicomponente para medir los tres componentes ortogonales de una fuerza y el par Mz. El umbral muy bajo y la alta sensibilidad permiten medir incluso las fuerzas más pequeñas. Con una frecuencia natural superior a 15 kHz en los tres ejes, pueden registrarse fuerzas muy dinámicas con una precisión extremadamente alta.
Carga
Carga