Capteur de force et de couple de réaction à 2 composantes, Fz max ±10 kN, Mz max ±25 Nm / 9345B

Capteur de force et de couple de réaction à 2 composantes, Fz max ±10 kN, Mz max ±25 Nm
9345B
18007692
  • Couplemètre préchargé et étalonné, prêt à mesurer immédiatement
  • Plages étalonnées : Fz : 0 ... +1 kN, 0 ... +10 kN ; Mz : 0 ... ±2,5 Nm, 0 ... ±25 Nm
  • Sensibilité : Fz : ≈ -3,7 pC/N, Mz : ≈ -190 pC/Nm
  • Dimensions : DxH (mm) 36x42
Sur demande
Sur demande
Measuring ranges
Minimum measuring range [kN]
-10 kN
Maximum measuring range [kN]
10 kN
Calibrations
Calibrated range Fz [kN]
0 … 10 kN
Product type
Force sensor type
A dynamometer is a force plate measuring forces and moments.
Multicomponent force / moment
Preloaded sensor
Preloaded sensors are already calibrated and therefore ready-to-use. Non-preloaded sensors are smaller.
Yes
Mechanical properties
Axial stiffness
1.7 kN/μm
Stiffness Cφ (Mz) [kN/μm]
0.19 N·m/μrad
Overload Fz [kN]
-12, 12 kN
Overload Mz [N·m]
-30, 30 N·m
Crosstalk
Maximum crosstalk Fx→Fz
-0.02 … 0.02 %
Maximum crosstalk Fy→Fz
-0.02 … 0.02 %
Maximum crosstalk Fx→Mz
-0.3 … 0.3 N·m/N
Maximum crosstalk Fy→Mz
-0.3 … 0.3 N·m/N
Maximum crosstalk Fz→Mz
-0.04 … 0.04 N·m/N
Maximum crosstalk Mz→Fz
-2 … 2 N/N·m
Maximum crosstalk Mb→Mz
-0.02 … 0.02 %
Maximum crosstalk Mb→Fz
-0.5 … 0.5 N/N·m
Metrological characteristics
Linearity (force)
Linearity represents the maximum deviation/error between ideal and actual output signal characteristics in relation to the measurand in a specific measuring range. It is expressed in percentage of the range of measurement signal (Full Scale Output).
≤±0.3 %FSO
Linearity (moment)
Linearity represents the maximum deviation/error between ideal and actual output signal characteristics in relation to the measurand in a specific measuring range. It is expressed in percentage of the range of measurement signal (Full Scale Output).
≤±0.3 %FSO
Hysteresis (force)
The maximum difference in output, at any measurand value within the specified range, when the value is approached first increasing and then decreasing measurand.
≤±0.5 %FSO
Hysteresis (moment)
The maximum difference in output, at any measurand value within the specified range, when the value is approached first increasing and then decreasing measurand.
≤0.5 %FSO
Nominal sensitivity Mz [pC/N·m]
-190 pC/N·m
Threshold Fz
Minimum change in the measurand that produces a measurable change in the sensor output, while the change of the measurand takes place slowly and monotonically.
0.02 N
Threshold Mz
Minimum change in the measurand that produces a measurable change in the sensor output, while the change of the measurand takes place slowly and monotonically.
0.02 mN·m
Electrical properties
Output signal type
Charge (Piezoelectric, PE)
Capacitance
≈340 pF
Minimal insulation resistance
10^13 Ω
Influence quantities
Sensitivity
≈–3.7 pC/N (Fz), ≈–190 pC/Nm (Mz)
Temperature coefficient of sensitivity Fz
–0.02 %/°C
Dynamical properties
Natural frequency Mz
Frequency of free (not forced) oscillations of the entire sensor. In practice the (usually lower) natural frequency of the entire mounted system governs the frequency behavior.
32 kHz
Certificates and Standards
Degree of protection EN 60529
IP65
Operation and installation
Operating temperature range
-40 … 120 °C
Tightening torque MA (recommended)
4 N·m
Applications
eVTOL / UAV / drone testing, 3C electronics manufacturing
Cable properties
Cable included
No cable included
Interfaces
Connector type
V3 negative
Dimensions and materials
Weight
267 g
Casing material
DIN 1.4542
Ce couplemètre piézoélectrique à 2 composantess est adapté à la mesure d'une force de traction ou de compression Fz et d'un couple Mz agissant sur le capteur. La haute résolution permet de mesurer les moindres variations dynamiques de forces et de couples importants sur des arbres non rotatifs. Des disques de quartz sensibles à la pression et au cisaillement sont intégrés sous forte précontrainte entre une vis de précontrainte creuse et un écrou qui constitue le boîtier du capteur. En raison de la rigidité extrêmement élevée des éléments du capteur, les composantes Fz et Mz sont mesurées pratiquement sans modification de la course ou de l'angle. Le signal de charge proportionnel à la force et au couple est capté par des électrodes et transmis par un câble blindé à l'amplificateur de charge. Ce dernier convertit les signaux de charge en tensions de sortie proportionnelles, qui peuvent être enregistrées et évaluées.
Chargement